| BiPhasic Waveforms and Energy
Levels
The Phillips Heartstart FR2 (formerly the Heartstream
Forerunner) was the first AED to use a Biphasic waveform, launched
back in 1996.
This device deviated from the Resuscitation Guidelines
of the time in two significant ways:
1. It dramatically lowered the energy setting to a
nominal fixed 150 joules for all shocks, compared to the established
200, 200, 360joule protocol.
2. It eliminated the established principle of escalating energy
settings for the third and subsequent shocks.
The potential benefits of Biphasic defibrillators
(which we all believe in), was the potential for improving the success
of defibrillation, whilst at the same time, being able to reduce
the amount of current delivered to the heart, thereby minimising
the possibility of damaging the myocardium.
In defibrillation however, we are faced with two challenges:
a. To deliver enough energy to effectively defibrillate
the Heart, bearing in mind that there are significant differences
across the patient population as to how much energy is required
to achieve this successful defibrillation.
b. To minimise the amount of current is passed through the Heart,
thereby minimising the chance of excess damage to the heart muscle.
In considering these challenges, I quote from the
International Resuscitation Guidelines, published in August 2000:
"Defibrillation depends on the successful selection
of energy to generate sufficient current flow through the heart
(transmyocardial current) to achieve defibrillation while at the
same time causing minimal electrical injury to the heart. A shock
will not terminate the arrhythmia if the energy and current are
to low. Selection of appropriate current also reduced the number
of repetitive shocks and limits myocardial damage".
"A more predictable increase in current occurs
when shock energy is increased. This supports second shocks at higher
energies".
It would seem appropriate, that when considering the
two challenges mentioned above, that we must favour systems that
ensure sufficient energy and current flow is achieved for all patients.
The PowerHeart AED with STAR Biphasic waveform therefore
is supported by these latest guidelines, as it can escalate energy
levels for subsequent shocks.
Indeed, PowerHeart and Survivalink FirstSave AEDs with STAR Biphasic
technology are the ONLY AEDs that can achieve this, by assessing
each patients individual characteristics and in response, varying
the energy level sufficiently to meet their needs.
By assessing each patient in this way, we also are
able to ensure that only sufficient energy is delivered, thereby
minimising the potential for myocardial damage.
As the Heartstart FR2 is a fixed low energy defibrillator
(i.e. there is no increase in energy available), then we must ask
how this can be achieved.
It is also worthy of note that all modern defibrillators
using a Biphasic waveforms introduced from 1999 onwards have employed
escalating energy protocols, and this includes other more recent
defibrillators introduced by Phillips / Laerdal.
It is interesting to note that only two devices were chosen for
the first phase of the 'Defibrillators in Public Places' initiative,
both units chosen were Biphasic (one being the Survivalink 'FirstSave
AED), however both were also set up to deliver escalating energies!
Clinical Data
In support of the above information, we are
able to clearly show clinical data that confirms our methodology,
whilst inevitably demonstrating the weakness of a fixed low energy
protocol.
| Age Of Technolgy |
PowerHeart AED |
Heartstart FR(2) |
| Product Released |
2002 |
1996 |
| |
|
|
| Hospital Data |
PowerHeart AED |
Heartstart FR(2) |
| Product Released |
100%4 |
87.1%5 |
| |
|
|
| Pre-Hospital Data |
FirstSave |
Heartstart FR(2) |
| Time to successful defibrillation |
~ 55 sec. |
~ 3 min. |
For every minute a heart is in sudden cardiac arrest,
the chance of survival decreases by 10%. The extra time it takes
to defibrillate 6 times versus 2.3 times reduces the chance of survival
for the patient by more than 20%.
1. Bain AC, et al., Annals of Emergency Medicine.
Accepted for publication, January 2001.
2. Bardy GH, etal., Circulation, 1996; 94:2507-2514.
3. Bain, et al., Pre-hospital Emergency Care, 2000; Vol14, No.1
(Updated).
4. Gliner BE, etal., Biomedical Instrumentation and Technology,
1998; 32:631-644.
5. Adgey AA, et al., American Heart Journal, 1988; 112:745-751.
6. Wilson, et al., European Heart Journal, 1998; 9:1257-1265.
7. Herlitz J., et al., British Heart Journal, 1994; 72:408-412.
8. American Heart Association, etal., Circulation, 2000; 102(supppl
I): I60-I76.
|